The infinity-corrected optical system reduces distortion and improves color accuracy.
Plan-achromatic lenses produce high-contrast images with edge-to-edge sharpness.
The Koehler illumination system provides centered, evenly-distributed lighting with a field aperture for maximum contrast and definition.
Rackless mechanical stage provides extremely smooth movement over a wide range.
Product Information
TECHNICAL SPECIFICATIONS
FRAME
Frame: Ergonomic design microscope with modular frame, on-site upgradable to Fluorescence, phase contras and darkfield microscopy. 6 position nosepiece to accommodate 6 objectives at a time
OPTICAL SYSTEM
Frame: Ergonomic design microscope with modular frame, on-site upgradable to Fluorescence, phase contras and darkfield microscopy. 6 position nosepiece to accommodate 6 objectives at a time
ILLUMINATION
The microscope should have an ergonomic stand with 12V/100W halogen light source or equivalent LED transmitted light source. Should have Light intensity manager switch with front knob for light intensity adjustment and built-in filters.
OBSERVATION
Should be Tilting trinocular wide field, with inclination from 5 degree to 30 degree three way light path distribution (100:0, 50:50 & 0:100) for simultaneous viewing and imaging of the specimens, provided with paired widefield eyepieces-FN 22 or more.
EYE-PIECES
Should be of at least IOX magnification, with diopter adjustment facility with field of view of 22 mm or higher.
NOSEPIECE
Reversed sextuple position revolving nosepiece to accommodate six objectives at a time with a slot for Analyser and Polarizer.
OBJECTIVES
Plan achromat objective 2X- 2.5X Plan achromat objective 4X Plan Apochromat objective I OX Plan Apochromat objective 20X Plan Apochromat objective 40X Plan Achromat Oil objective I OOX Oil
STAGE
X and Y rectangular mechanical stage motion control on right hand side with capacity to hold two slide glasses at a time. The stage should be ceramic coated and rotatable.
REVOLVING NOSEPIECE
Revolving nosepiece to accommodate 6 objectives at a time with a slot for analyser and polarizer. Separate slot for tint plate, for urate crystals and CPPD crystals should be present. Tint plate should also be provided.
CONDENSER
swing out. Condenser should be compatible with 2X-1 OOX and also compatible with achromat and apochromat objectives
MULTI-HEAD ATTACHMENT
5 additional ergonomic viewing heads apart from the primary user. In all viewing end illumination should be as strong as in the master objectives. Both eye pieces should have diopter adjustments for all secondary observers and should have FOV 22mm and above. When using the same type of observation tube, orientations of images observed by all observers should be identical.
LED ARROW POINTER
Multicolor (minimum 2), intensity controllable LED Colored indicator should be supplied with the microscopes. Arrow control Joystick, arrow rotation dial and color selection dials should be provided.